5The Pulse Oximeter can be used in measuring the pulse oxygen saturation, pulse rate and perfusion index through finger.The product is fit for family
63 Principle Principle of the Oximeter is as follows: An experience formula of data process is established taking use of Lambert
74 Technical specifications 4.1 Main performance SpO2 value display Pulse rate value display, bar graph display Pulse waveform dis
8C Perfusion Index Range: 0%~20% D Resolution SpO2 : 1%, Pulse rate: 1bpm; Perfusion Index:0.1% E Measurement Performance in Weak Filling
9Figure 2 Front View 5.2 Probe installation A. Open the USB cover,inserting the probe into the USB port of the pulse oximeter,and the red-ligh
10 Figure 3 Fingernails and the luminescent tube should be in the same side. 6.1.3 Pause Alarm a. Alarm includes the alarm of measure
11A Brightness Adjustment In the main menu interface,short press menu button to move the choice bar to the "Brightness" item,then lo
12Figure 5 Alarm Setting Menu a The high/low limit of alarm setting In alarm setting menu,short press menu button to move the choice bar to
13setting menu as Figure 6. . Figure 6 Record setting Menu a Record segments Display the number of the data segments which have been recorded.
14In the main menu interface,short press menu button to move the choice bar to the "System" item,then long press menu button to en
IIInstructions to User Dear users, thank you very much for purchasing the Pulse Oximeter. This Manual is written and compiled in accordanc
15transmission.If the user can't set "Wireless" after cutting connecting,please try again after waiting 30 seconds. Recommended adapt
16to "Exit” item, then long press button to exit main Menu and return to the measure interface. H. Clock Setting In the measure interface, S
17 When charging in power-off state, short press power button and the device will display green dynamic state bar.It means that the device
18B.The finger should be in a proper position (see the attached illustration as Figure 3 for reference), or else it may result in inaccurate measure.
19may also be a major factor resulted in serious error of SpO2 measure. D. The SpO2 value serves as a reference value for judgment of anem
208 Troubleshooting Trouble Possible Reason Solution The SpO2 and Pulse Rate can not be displayed normally. The finger is not properly p
21The device can not be used for full time after charge The battery is not full charged. Please recharge the battery The battery is broken
229 Key of symbols Signal Description Warning – See User Manual %SpO2 The pulse oxygen saturation (%) PRbpm Pulse rate (bpm) PI Perfusion I
23 Menu button power button IP22 International Protection USB WEEE (2002/96/EC)
2410 Function specification Information Display Mode The Pulse Oxygen Saturation(SpO2) 2-digit digital OLED display Pulse Rate(PR) 3-digit
IIIthan 2 hours. For the special patients, there should be a more prudent inspecting in the placing process. The device can not be clippe
25Battery Requirement Voltage 3.7 rechargeable lithium battery × 1 (The red wire on the battery denotes anode,the black wire on the batter
26Low voltage alarm 60s 5ms SpO2 alarm 1s 5ms Pulse rate alarm 1s 5ms Probe error alarm 16ms 5ms
IV1.1 Instructions for safe operations ... 1 1.2 Warning ...
11 Safety 1.1 Instructions for safe operations Check the main unit and all accessories periodically to make sure that there is no visi
2 Please don't measure this device with functional tester for the device's related information. 1.3 Attention ֠ Keep the oxi
3֠ If some abnormal conditions appear on the screen during test process, pull the finger out and reinsert to restore normal use. ֠ The dev
42 Overview The pulse oxygen saturation is the percentage of HbO2 in the total Hb in the blood, so-called the O2 concentration in the bloo